DESKTOP/TROLLY PATIENT MONITOR
WITH ECG ANALYSED FROM PULSE |
|
 |
 |
Compact, light, easy to carry and use |
 |
Large, bright, easy-to-read LED and 4" LCD displays |
 |
Wireless or network connection to patient-monitoring system |
 |
Storage and review of NIBP and SpO2 data - up to 1200 groups |
 |
3-level audio and visual alarms |
 |
Adult, paediatric and neonatel programs |
 |
Monitors NIBP, Sp02, pulse rate and temperature |
 |
Removable, rechargeable battery |
|
|
|
 |
|
|
|
TECHNICAL SPECIFICATION |
| Power requirements |
| INTERNAL BATTERY |
Type |
Lead acid, removable 12V 2AH |
Operating time |
3 hours |
Recharge time |
6 hours |
AC power |
100-240 V 47-63 Hz 70VA |
Alternative DC input |
10-15 V DC |
| NIBP |
Measurement method |
Oscillometric |
Measurement unit |
mmHg/kPa selectable |
| Measurement range |
Adult |
10-270 mmHg |
Paediatric |
10-235 mmHg |
Neonatal |
10 -135 mmHg |
Measurement mode |
Manual, auto, continuous |
Measuring interval
in AUTO |
1,2,3,4,5,10,15,20,30,45,60,90 min, 2, 4, 8, 12 hours |
Measuring period in STAT |
5 min |
Over-pressure
protection |
Yes |
| SpO2 |
Measurement range |
0-100% |
Accuracy |
±2% (70-100%), ±3% (35-69%),
0-34% unspecified |
Resolution |
1% |
| Pulse rate |
Measurement range |
25-250 bpm |
Accuracy |
±1 bpm |
Resolution |
1 bpm |
| TEMP |
Range |
0 -50˚C |
Resolution |
1˚C |
Accuracy |
±0.1˚C (25 - 50˚C) |
| Alarm |
Mode |
Audio/visual |
Type |
SYS, DIA, MEAN, TEMP, PR, SpO2, all 3 levels |
| Storage and review |
1200 groups of NIBP, 48 hours trend data. |
| Displays |
4" LCD and 7 segment LEDs
|
OPTIONAL THERMAL PRINTER |
| Dimensions |
256 x 209 x 170 mm |
| Net weight |
3.4kg with battery |
|
|
|