N104 Daray Ltd
N104
N104
BSI
 
Nfuze title N104 title
VOLUMETRIC INFUSION PUMP
The N104 volumetric infusion pump is a digitally controlled device for use in most types of intravenous/arterial infusion, providing a reliable alternative for fluid control especially for long-term, micro-fluid and precision injections.
The DARAY N104 is designed for use for intravenous or arterial infusion in medical treatment and nursing throughout many different sectors.
N104
Take control
The rate of infusion is automatically controlled with no effect from the method of intravenous or arterial method, reservoir height or medicine viscosity.
The delivery volume is adjustable from 1 to 9999ml with automatic switching to KVO
(keep vein open) rate. Dosage calibration can be adjusted when necessary using the DIP switches inside.
Internal battery
The N104 pump operates from an internal battery with a built-in automatic charger, which charges the battery whenever the unit is connected to the mains. The battery also powers the unit in the event of mains failure (at a rate of 50ml/h for up to 3 hours). Alternatively, the unit can be powered from an external 12V supply.
Remote alarms
The alarm functions can be sent to a supervisor’s desk or monitoring station using the ‘alarm signal socket’ on the back of pump. The alarm buzzer can be silenced using the ‘SELECT’ button.

Anaesthesia
Radiography
Emergency room
Paediatrics
Cancer therapy
General surgery
Heart surgery
Thoracic Surgery
Peritoneal procedures
Obstetrics
Gynaecology
Extensive alarms
Complete
When the infusion dosage has reached its limit, the pump stops automatically and enters KVO state with a rate of 1 ml/h.
Air-in-line
Air in the line stops the pump automatically and enters KVO state.
Occlusion
If the infusion tube is pinched or the device is blocked, the pump stops automatically
Complete/Occlusion
The infusion fluid is finished. The pump stops automatically and enters KVO state
Door open
The door has been opened during infusion. The pump stops automatically.
Low battery
The internal battery charge is low. The pump should be connected to 230V AC mains as soon as possible.
The alarm clears after the battery has been charging for several minutes.
N104 attached to drip stand
Accessories
Supplied with the N104:
Drip-stand clamp
Drip chamber detector
20 sterilized IV sets
Spare fuse
2 pin signal plug
4 pin signal plug
Power cable
User manual

TECHNICAL SPECIFICATION
Flow range
1 to 9999 ml
Flow rate selection
1 to 1000 ml/h
Volume limit
1 to 9999 ml
Infusion accuracy
1 to 1000 ml ±10%
Dimensions
180 x 125 x 260 (mm)
Weight
4.25kg