N103 syringe Pump daray logo  
 
N103 syringe Pump


N103 syringe Pump product brochure

 
N103 syringe pump
N103 syringe pump
   
  The N103 syringe Pump is a digitally controlled device for use in most types of intravenous/arterial infusion.
 
Accuracy and control
The N103 has a simple and clear display and controls protected by a waterproof membrane. The device automatically activates self-test, self-diagnosis, and error alert procedures each time it is powered on.
Rapid purge
Whilst in ‘HOLD’ mode the driver can be run at high speed to either expel the air from the syringe or to purge the syringe. The rate of purge depends on the size of the syringe fitted.
 
Dosage calibration
DIP switches on the underside of the device are provided for calibration to cater for variance between syringes. Choose from -5%, STD (default), +5%, +10% and +20%.
'CHECK' function
Pressing the `CHECK´ button displays the volume of medication delivered at any time whilst the device is running.

Extensive alarms
Near empty
When there is approximately 5ml of medication left, an audible and visual alarm is activated.

Complete
When infusion is complete, the operation stops and audible and visual alarms are activated.

 
Occlusion
If occlusion occurs, operation stops. The MUTE button silences the alarm but infusion can only continue when the alarm condition is cleared.
 
Low battery
If the battery runs low, audible and visual alarms are activated. The pump can continue to operate whilst recharging.
 
Automatic syringe size detection
The N103 automatically detects the size of syringe and controls the flow rate rather than simply regulating the infusion time.
 
Powerful choices
The N103 syringe pump operates from an internal battery which is charged when the unit is connected to the mains and powers the unit in the event of mains failure (up to 4 hours with maximum charge at a flow rate of 50ml/hr). Alternatively, the unit can be fed from an external 12V supply.
Rate of purge
Syringe size (ml)
flow rate (ml/hr)
20
150
50
300
100
400
Technical specification
Infusion accuracy
± 5% (including syringe accuracy)
Acceptable syringe sizes
20ml, 50ml, 100ml
Flow-rate ranges
0.1ml/hr - 300ml
(0.1ml/hr increments)
Operating temperature
5 - 40°C
Relative humidity
80% max.
Power requirements
220V AC 50Hz 25W or 12V DC 25W
Battery duration
4 hrs (at 50 ml/hr flow rate)
Dimensions
335×142×120 mm
Weight
2.5 kg
 


Features

Digital control
Dosage calibration setting
CHECK function
Rapid purge
Occlusion detection
Automatic syringe size detection
Extensive alarms
Power-on self-diagnosis