Dosage calibration
DIP switches on the underside of the device are provided for calibration to cater for variance between syringes. Choose from -5%, STD (default),
+5%, +10% and +20%. |
'CHECK' function
Pressing the `CHECK´ button displays the volume of medication delivered at any time whilst the device is running. |
Extensive alarms
Near empty
When there is approximately 5ml of medication left, an audible and visual alarm is activated. |
Complete
When infusion is complete, the operation stops and audible and visual alarms are activated. |
|
TECHNICAL SPECIFICATION |
Infusion accuracy
(inc. syringe accuracy) |
± 5% |
Syringe sizes |
20ml, 50ml, 100ml |
Flow-rate ranges |
0.1ml/hr - 300ml
(0.1ml/hr increments) |
Operating temperature |
5 - 40°C |
Relative humidity |
80% max. |
Power requirements |
220V AC 50Hz 25W or 12V DC 25W |
Battery duration |
4 hrs (at 50 ml/hr flow rate) |
Dimensions |
335×142×120 mm |
Weight |
2.5 kg |
|
Occlusion
If occlusion occurs, operation stops. The MUTE button silences the alarm but infusion can only continue when the alarm condition is cleared. |
Low battery
If the battery runs low, audible and visual alarms are activated. The pump can continue to operate whilst recharging. |
|
Automatic syringe size detection
The N103 automatically detects the size of syringe and controls the flow rate rather than simply regulating the infusion time. |
Powerful choices
The N103 syringe pump operates from an internal battery which is charged when the unit is connected to the mains and powers the unit in the event of mains failure (up to 4 hours with maximum charge at a flow rate of 50ml/hr). Alternatively, the unit can be fed from an external 12V supply. |
|