Full patient monitor
(portable size) |
|
 |
|
|
TECHNICAL SPECIFICATION |
Safety |
IEC60601 (GB9706)
|
Dimension |
256mm (W) x 209mm (H) x 185mm (D)
|
Weight |
4 Kg
|
Power supply |
100-240V AC47- 63 Hz |
Alarm indication |
Visual and audible alarm
|
| ECG |
Leads |
I, II, III, aVR, aVL, aVF, Cx |
Gains |
2.5, 5, 10, 20, 40mm/mV and auto |
Input impedance |
≥5MΩ |
Common mode rejection ratio |
≥85dB (diagnostic mode)
≥105dB (monitoring and surgical mode) |
Bandwith |
0.5-40Hz (monitoring mode)
0.05-120Hz (diagnostic mode)
1-25Hz (surgical mode) |
Electrode disconnection detection |
Yes |
Cardiogram playback |
Yes |
Scanning speed |
12.5mm/s, 25mm/s, 50mm/s |
Heart rate range |
10-300 BPM |
Accuracy |
±0 BPM |
| NIBP |
Method |
Automatic oscillometric |
Operation modes |
SINGLE (manual), CYCLE (automatic),
RAPID (continuous)
|
Measurement |
Systolic pressure/diastolic pressure/mean |
Range: Adult |
10-270mmHg |
Pediatric |
10-235mmHg |
Neonatal |
10-135mmHg |
Unit |
mmHg / kPa selectable |
Auto Measurement |
Minutes: 1, 2, 3, 4, 5, 10, 15, 20, 30, 45, 60, 90
Hours: 2, 4, 8,12 |
| 600 groups of NIBP measuring data review |
| RESP |
Method |
Impedance measurement
|
Measuring range |
0-120 BPM (180 BPM max) |
Alarm limit |
0-120 BPM, upper and lower limit continuously adjustable |
Apnea alarm |
Yes |
| BLI DIGITAL SPO2 (PULSE OXIMETRY) |
Range |
0-100% |
Accuracy |
±2% (70-100%);
±3% (35- 69%)
Unspecified (0-34%) |
Resolution |
1% |
| TEMP |
Channel |
2 channels |
Range |
0-50°C |
Unit |
Celsius, Fahrenheit |
Resolution |
1mmHg |
Accuracy |
±0.1°C:(25 - 50°C) |
| PULSE RATE |
Range |
10-300 BPM |
Accuracy |
±1 BPM |
Resolution |
1 BPM |
| ENVIRONMENTAL |
Temperature |
Operation 0-50°C
Storage/transportation: -40-70°C
|
Humidity |
Operation: 10%-95%, noncondensing
Storage/transportation: 0%-93% noncondensing
|
Atmospheric
pressure: |
Operation: 700-1060hPa,
Storage/transportation: 500-1060h Pa |
|
|
|
| |
 |
|